OPERATING PROCEDURE:

ACCESSORIES

Upon receiving the shipment, inspect the container and equipment for any signs of damage.
Verify that you have received all the items listed below:
Meter, Conductivity cell, Buffer Solution pH 4 & 7, Standard Solution 1413 S/cm, Soaking
Solution, Lanyard, Battery (installed), Instruction manual, Carrying case.
Optional: ORP electrode

PREPARATION

1. Remove the protective cap and unscrew soaking bottle #(pH, ORP only) from meter. Rinse the

electrode with clean water and wipe it dry. Don't leak soaking solution from bottle and place
the protective cap back on the bottle at the end of usage.

2. Press

button to turn the meter power on.

Note: Don’t touch or wipe the surface of the inner black sensor of the conductivity cell.

CALIBRATION
pH

1. Make sure the sensor is pH electrode or check the icon

shown on the LCD display.

2. Dip the electrode into the buffer solution pH 7. Stir gently and wait until the display is

stabilized. Press and hold

button to enter calibration mode until the icon

CAL appears on

the display, and # flashes 7.00. When the display stops flashing and indicates “SA”. It will be
followed by “End” while calibration is ending, and then will return to measurement mode.

3. Rinse the electrode with clean water and wipe it dry. Dip the electrode into the buffer

solution pH 4. Stir gently and wait until the display stabilizes. Press and hold

button to

enter calibration mode until the icon

CAL appears on the display, and flashes 4.00. When the

display stops flashing and indicates “%” (percentage of slope), then “SA”. It will then be
followed by “End” while calibration is ending, and then return to measurement mode.

4. After slope calibration pH 4 or pH 10, the display will indicate percentage of slope (PTS) to

show the status of electrode. If the PTS is below 70% or above 130%, the electrode must be
replaced. A slope of 100% is ideal.

Note: (1) Icon “SA” will not appear if the calibration fails.

(2) When doing a 2 or 3 point calibration, Calibrate with buffer pH 7 first, and then follow

with buffer pH 4 or pH 10.