beautypg.com

Specifications – General Tools and Instruments UV340 User Manual

Page 5

background image

3. SPECIFICATIONS

Display

LCD size : 18 mm x 45 mm.
Maximum indication 1999.

Measurement

2 ranges : Hi & Lo range

ranges &

Lo range : 1999 uW/cm^2 x 1 uW/cm^2

resolution

Hi range : 19990 uW/cm^2 x 10 uW/cm^2

* 1000 uW/cm^2 = 1 mW/cm^2

UV sensor

Band pass 290 nm to 390 nm.

spectrum
Accuracy

± ( 4 % FS + 2 dgt )

FS : full scale

* Calibration is executed under the UVA light &

and compare with the standard UVA light meter.

* Spec. tested under the environment RF Field

Strength less than 3 V/M & frequency less than the
30 MHz only.

Sensor structure

The exclusive UV photo sensor with
the cosine correction filter.

Buttons. Switch

Power button, Zero button, Range switch

Sample Time

Approx. 1 sec.

Over Range

Indication of " ".

indication
Weight

251 g / 0.55 LB (including battery)

Operating Temp.

0 to 50 .

Operating

Less than 85% R.H.

Humidity
Power Supply

DC 9V battery, 006P , MN 1604 (PP3)
or equivalent.

Power

Approx. DC 6 mA.

Consumption
Size

Main instrument : 131 x 70 x 25 mm.
Sensor probe head : 45 mm dia x 32 mm.
Sensor probe handle : 125 x 24 mm dia.

Accessories

Instruction manual........... 1 PC.

Included

UV sensor probe.............. 1 PC.

Optional Acc.

Soft carrying case, CA-52A, CA-03.

2