beautypg.com

General Tools and Instruments DMM51FSG User Manual

Page 4

background image

B-2

ALTERNATING CURRENT VOLTAGE (AC VOLTAGE)

Range

Resolution

Accuracy

DMM51FSG

200mV

100

␮V

± (1.2% rdg + 3 digits)

2V

1mV

20V

10mV

± (0.8% rdg + 3 digits)

200V

100mV

750V

1V

± (1.2% rdg + 3 digits)

INPUT IMPEDANCE: All ranges are 10M

Ω.

FREQUENCY: 40Hz-400Hz.
OVERLOAD PROTECTION: 200mV is 250VDC or AC RMS.

All other ranges is 750Vrms or 1000Vp-p.

DISPLAY: Average Value (RMS of Sine Wave).

B-3

DIRECT CURRENT CURRENT (DC CURRENT)

Range

Resolution

Accuracy

DMM51FSG

20

␮A

0.01

␮A

± (2% rdg +5 digits)

200

␮A

0.1

␮ A

± (0.8%rdg +1 digit)

2mA

1

␮A

± (0.8% rdg +1 digit)

20mA

10

␮A

200mA

100

␮A

± (1.5% rdg +1 digit)

2A

1mA

± (1.5%rdg +1 digit)

10A

10mA

± (2%rdg +5 digits)

OVERLOAD PROTECTION: 2A/250V fuse (below 2A range) and 10A/250V fuse

(10A range).

MAX CURRENT INPUT: 10A
MEASURING VOLTAGE DROP: Full range is 200mV.

B-4

ALTERNATING CURRENT CURRENT (AC CURRENT)

Range

Resolution

Accuracy

DMM51FSG

200

␮A

0.1

␮A

± (1.8% rdg +3 digits)

2mA

1

␮A

± (1% rdg +3 digits)

20mA

10

␮ A

± (1% rdg +3 digits)

200mA

100

␮A

± (1.8% rdg +3 digits)

2A

1mA

± (1.8% rdg +3 digits)

10A

10mA

± (3% rdg +7 digits)

OVERLOAD PROTECTION: 2A/250V fuse (below 2A range) and 10A/250V fuse

(10A range).

MAX CURRENT INPUT: 10A
MEASURING VOLTAGE DROP: Full range is 200mV.
DISPLAY: Average Value (RMS of Sine Wave ).

DMM51FSG Manual FINAL-091908:Layout 1 9/19/08 4:03 PM Page 4