Section 6: Parts

11/12

54

51267 Rev. 2

Replacing an RC load cell

1.

Remove power from the instrument.

2.

Lift the scale using a proper size and rated hydraulic jack (or jacks) at the corners
closest to the defective load cell location.

3.

Check upper and lower receiving cups and “O” rings for damage. Replace as
necessary and reapply grease.

4.

Insert the new cell into the upper receiving cup.

5.

Carefully lower the hydraulic jack until the cell is set into the lower cup.

6.

Remove the cover of the J-box, then loosen the gland bushing to free the cable.
Remove the old cell wires and connect new cell wires in the balance box/SSC.
Secure the cover.

7.

Test and adjust the scale as necessary.