Eliminate mechanical constraints – The vessel should be fully supported on
the Omnicells. Reduce or eliminate any mechanical issues which will interfere with
the vessel’s deflection, such as: rigid vertical or horizontal piping (use flexible
piping when connecting to the vessel whenever possible), ladders, catwalks, etc.
Orientation – Orientation of the 9120 Series Omnicell isn’t critical because of the
rocker load cell design, however it is recommended positioning the load cell
cables away from traffic areas.
Heated or chilled vessels –
Consult with the appropriate Product Manager for
any application with vessels that heat or chill the product inside the vessel.
In these cases, special consideration may be required to isolate the temperatures.
Load cell cabling – The load cell cables should be routed away from areas
where they could be damaged. Use drip loops at both the load cell and the
junction box ends.
Hazardous area – If the Omnicells will be installed in a Hazardous Area, a
document is required from the end user stating the Class, Division, and Group.