Section 4: Wiring

11/12

34

51182 Rev. 2

DATA RECORDING

1. Record scale serial number from the tag.

2. Record instrument, junction box and load cell serial numbers.

3. Keep a copy of the sheet in the customer file.

MOISTURE PROTECTION

Full Electronic Scales have been designed to provide protection from the effects of
moisture.

• The load cells are calibrated with the cable attached.
The cable

MUST NOT

be cut.

• The cable is connected directly to the Junction Box through a sealed bushing

which

MUST BE TIGHTENED WITH PLIERS

to keep moisture out of the box.

• On all boxes, the black plastic fittings have O-rings that can be forced out of

position if the bushing itself is not tight.

To prevent this, first tighten the inner nut securing the bushing in the hole.

Then insert the cable and carefully tighten gland with pliers until it is very snug.

DO NOT over-tighten where bushing 'turns'.

All box covers MUST BE SECURED with ALL screws tightened properly (

18

to 20 lbs/in

) for protection against moisture.