Appendix AX300

Teledyne

Analytical

Instruments

28

Appendix

A.1 Specifications

Range: 0-100% oxygen

Accuracy: +2% of full scale (at constant temperature

and pressure)

Response

Time: 90% in less than 6 seconds at 25 °C

Battery

Life: Approximately 2000 hours of continuous

use in non-alarm condition

System

Power:

3 AA alkaline batteries.

Sensor

Type:

Class R17MED

Alarm: Non-adjustable

 SENSOR alarm

activates upon sensor problem or O

2

con-

centration falls below 18%

Expected Life: 36 months in air at 25°C, Atmospheric

Pressure and 50% Relative Humidity.
(~7.5 months under similar conditions
when continuously exposed to 100% oxy-
gen)

Dimensions: 2.5" W × 1.25" D × 4.5" H (66 mm x 33-

mm × 111.5 mm)

Sensor

Cable: Retracted: 2 ft / Extended: 10 ft.

Storage Temp. 10-30°C (continuous), 5-50°C (Intermit-

tent)

Operating Temp: 0-40°C