Portable Oxygen Analyzer

Operating Instructions

Teledyne

Analytical

Instruments

13

When properly calibrated, the output signal generated by the analyzer is li-

near and proportional to the oxygen concentration.

If you requested Option-B (RS 232 digital output) at the time of purchase, a

digital RS 232 signal is output from the output port shown in Figure 2-8. Use the
optional RS 232 Interface Cable (P/N B-75555) available from Teledyne for
connection to a standard RS 232 port on a computer or other suitably equipped
digital device.

CAUTION:

AX300 ANALYZER SHOULD ONLY BE CONNECTED TO AN
EN60601-1/IEC60606-1 APPROVED DEVICE.

Figure 2-8: 0-1 VDC or RS 232 Digital Output Port

RS232 signal output has been tested with Windows HyperTerminal. The setup

changes with different windows versions. If other software other than windows
HyperTerminal is used, consult with factory.

Windows 95 version

Baud: 2400
Data:

7

Parity: ODD
Stop bit:

1

Flow control:

NONE