2.3.6 Cleaning

The AX300 and R17MED sensor are non-sterile devices.
The oxygen analyzer, oxygen sensor, and sensor interconnection cable may

be cleaned by wiping the surfaces with isopropyl alcohol or a mild cleaning solu-
tion.

When cleaning, do not allow liquids or moisture to enter the analyzer or

sensor internal cavities. Do not expose the electrical connections to the cleaning
solution. Do not immerse the analyzer, oxygen sensor or sensor interconnection
cable in water or any other liquid.

Do not expose the analyzer, oxygen sensor and interconnection cable to

steam, ethylene oxide, or radiation sterilization.

After cleaning, ensure that all surfaces are dry and the analyzer is properly

calibrated before using the analyzer.

2.4 Dos and Don’ts

– DO –

Read and understand the manual before using the analyzer

Calibrate every 8 hours or before every use.

Visually inspect the sensor for leakage before each use.

Calibrate using 100% oxygen and check in air.

Test batteries regularly and replace when battery indicator shows low
battery (no bars)

Keep the analyzer, sensor and connections dry, or on the dry side of
the breathing system.

Recalibrate after replacing the batteries.

Recalibrate after replacing the sensor.

Use only alkaline batteries.

Ensure that the R17MED sensor is properly connected to the analyz-
er.

Always use the plastic flow diverter when using the tee adapter.

Remove and save the plastic flow diverter when using the sensor in
non-flowing applications

Use only isopropyl alcohol or mild detergent, if cleaning is required,