Portable Oxygen Monitor

Service Manual

Teledyne Analytical Instruments

35

Table 3-2 Error Codes

Error Code

Audio Beeps

Error

Indeterminate

2

The watchdog timer has timed out indicating a
serious software error

30

3

Analog output is different from the expected value.
May indicate a shorted or over-loaded analog
output or a failure of the analog to digital converter
or digital to analog converter circuit.

50

5

The ADC circuit failed during POST.

6 (5)

6

(There is no
indication of
which key is
stuck)

A stuck key has been detected. The second digit,
in this example (5), on the display shows which
key is stuck:

0 - Low Alarm

1 - Batt

2 - Cal

3 - Up

4 - Down

5 - Silence

6 - High Alarm

7 - Alarm Test

8 - Key Lock

7(2)

7

(There is no
indication of
the type of
DAC failure)

A failure has occurred during the automatic
calibration of the digital to analog converter (DAC)
circuit. The second digit, in this example (2),
shows the type of digital to analog failure.

0 - Measurement

1 - High Test

2 - Low Test

3 - Offset Cal

4 - Gain Cal