Portable Oxygen Monitor

Operation

Teledyne Analytical Instruments

21

5. To disable the alarms set the high alarm set point above 100%.

Press LOCK/UNLOCK key to save settings.

Note: When the alarm is in the OFF condition (set point above

100%) the ALARM OFF status message blinks slowly on
the LCD below the oxygen readout.

2.1.6 Output 0-1 VDC or RS232

Both the AD300 and MD300 provide signal outputs for use with

recorders and computers. The instruments are supplied standard with a
0-1 VDC output. An optional 0-1 VDC Interface Cable (P/N B-75554)
is available from Teledyne for this purpose.

To connect the analyzer to an analog recording device:

1. Insert one end of the interface cable into the output port on

the side of the instrument. See Figure 2-9.

2. Insert the other end into the analog recorder device. Make

sure the device is equipped to handle a 0-1 VDC signal.

When properly calibrated, the output signal generated by the

analyzer is linear and proportional to the oxygen concentration.

If you requested Option-B (RS 232 digital output) at the time of

purchase, a digital RS 232 signal is output from the output port shown in
Figure 2-9. Use the optional RS 232 Interface Cable (P/N B-75555)
available from Teledyne for connection to a standard RS 232 port on a
computer or other suitably equipped digital device.