Figure 2-4: V-Mount Adapter Installation

2.1.2.2

U

NIVERSAL

M

OUNTING

C

LAMP

I

NSTALLATION

The Universal Mounting Clamp is supplied with a 1/4-20 screw for

securing the clamp to the rear of the instrument. A threaded brass insert is
installed on the back of the instrument for this purpose. See Figure 2-5.

Figure 2-5: Brass Insert for Universal Mounting Clamp

2.1.3 Battery Installation

Note:

Three “AA” alkaline batteries must be installed in the unit
before the monitor will operate. The unit must be
recalibrated whenever new batteries are installed
. For
the MD300, the HI and LOW alarm set points must be
reset to the desired values.

To install the batteries:

1. Turn the unit off (if it is on).

2. Hold the instrument face down in the palm of your hand. Use

a coin to pry up the latch that secures the battery
compartment door. Remove the battery compartment door.