beautypg.com

Operation, 1 setup, 1 sensor installation or replacement – Teledyne MD300 - Oxygen analyzer and monitor (Diving) User Manual

Page 13

background image

Portable Oxygen Monitor

Operation

Teledyne Analytical Instruments

13

Operation

Note: Upon receipt, INSPECT THE ENTIRE UNIT FOR

DAMAGE. Check the unit and all included accessories for
broken or loose parts. If damaged, DO NOT USE. Notify
the shipper, and consult
Teledyne Analytical Instruments.

Note: This equipment is internally powered using 3 AA batteries.

2.1 Setup

To set up and use your AD300 analyzer or MD300 monitor:

1. Install the sensor.

2. Install the batteries.

3. Calibrate the unit.

4. Set the alarms (MD300 only).

The control keys are designed for easy operation. A LOCK/UNLOCK

key has been incorporated to prevent accidental changes to critical settings.
This eliminates unwanted changes in calibration or alarm settings from
accidental touching or bumping of the keys. To further reduce the possibly of
incorrect adjustments at least two keys must be pressed in order to modify a
critical calibration or alarm set point value.

Note: The ALARM SILENCE () and BATT TEST key continue

to operate normally when the lock feature is activated.

2.1.1 Sensor Installation or Replacement

Note: The R17S oxygen sensor must be installed before the

oxygen analyzer/monitor can be operated

1. Remove the new sensor from its protective bag. Inspect the

sensor for damage or electrolyte leakage. If the sensor is
damaged, obtain a replacement. Do not use the defective sensor
as it may damage the unit.

This manual is related to the following products: