beautypg.com

Teledyne 8800A/8800B - Panel / Wall mount trace moisture analyzer User Manual

Page 58

background image

8800 Series Trace Moisture Analyzer Instruction Manual Appendices

51

Relationship of 4-20mA output and instrument reading in ppmV or LBS

Instrument reading in ppmV

Instrument reading in LBS of H2O / million cubic feet

10

100

1000

10000

4.1mA

5mA

0.1

1

10

100

1000

12mA

8mA

20mA

4.5mA

4/20 mA output

0.56 LBS = -60˚C = 10.66 ppmV

1.97 LBS = -50˚C = 38.83 ppmV

49.3 LBS = -20˚C = 1019.3 ppmV

289.1 LBS = 0˚C = 6063.8 ppmV

Low

AO range = -100˚C High

AO range = +20˚C

Low

AO range = -65.5˚C

High

AO range = -38.5˚C