8800 Series Trace Moisture Analyzer Instruction Manual

30

Legend: + denotes a beeping accompaniment to the message.

/

denotes alternately flashing messages.

DISPLAY

EXPLANATION

REQUIRED ACTION

LO ERR

+

PROM check sum failed.

power

on tes

ts

HI ERR

+

RAM write/read test failed.

cycle power

% ERR

+

Unidentified power-up failure.

if problem persists,

C.S.F/

+

EEPROM Check Sum Failed.

return to your representative

ERR/AD

+

A/D converter failure.

sy

st

em

te

st

s

on

ce p

er 2

m

in

.

for service.

ERR/REF

+

Reference voltage for A/D out of spec.

LO/BAT

+

Low supply voltage.

LO RNG/

+

TEMP

Instrument low temperature range

has been exceeded.

make sure that the unit is at a

HI RNG/

+

TEMP

Instrument high temperature range

has been exceeded.

temperature of -10°C to +50°C.

OPN/

+

Sensor circuit is open.

dewpoint dis

play

te

st

s

SAT/

+

Sensor is saturated.

see table in sensor troubleshooting section 2.5

SHR/

+

Sensor circuit is shorted.

...

Trying to calculate dewpoint for undefined sensor.

select sensor and recalibrate.

DB

+

Trying to calibrate an undefined sensor.

cal

ib

ra

tio

n

te

st

s

see calibration instructions.

SEN/2LO

+

Sensor reading is ‘too’ low to be from a saturated sensor,

thus can not be calibrated

2LO

While Compensating for cable length, the measurement did not

correspond to the expected open or short values.

cabl

e

te

st

s

Check the means of opening & shorting the sensor

cable. see section 3.4.5 - 11

2HI

ERR/EEP

+

EEPROM write cycle not completed.

m

is

cel

laneous

if this persists, return for service.

LOC

+

Attempting to modify a locked unit.

unlock unit, see set-up state 3.4.5.-10.

SEC

+

Attempting to unlock a secured unit.

HI (flashing while
viewing DEWPOINT)

The HI Alarm (alarm #1) relay is deenergized.

See section 3.4.4.2 Setting the Alarms.

LO (flashing while
viewing DEWPOINT)

The LO Alarm (alarm #2) relay is deenergized.

A.O./OPN

+

The Analog Output ‘sees’ an open circuit

check the cables connecting the analog output, make

sure the terminating impedance is less that 500

Ω.

PSI flashing when dis-
playing dewpoint

Pressure correction is enabled, the dewpoint is computed using

programmed pressures

refer to section 3.4.4.1

-70

°C (-94°F) with

flashing DEWPOINT

An XTR-65 sensor is measuring less than -70°C or -94°F D.P.

check instrument

upgrade to XTR-100 sensor

RNG

number can not be displayed in 3½ digits

XEN

+

Turn on message, Xen (Greek

Ξ=X)

display is dark

No power to instrument

check power to instrument, check fuse