Infrared Gas Analyzer

Maintenance 5

5-3

Teledyne Analytical Instruments

cells can be swabbed with water and alcohol and dried thoroughly before
reassembly. Sealed cells can be blown out with air, washed, or replaced.

C) After the cells have been cleaned and thoroughly dried,

reassemble them into the optical bench. Also attach any disconnected wiring
back onto the bench and check all connections and fasteners.

D) Turn on the analyzer and allow it to warmÄup for at least ten

minutes. Check for zero and span gas response.

NOTE: Typically, sample cell should only be removed for cleaning

when the Auto Zero circuit can no longer com~ensate for offset or if there is
a loss of sensitivity caused by cell lnefficiency .

5.4

System Self Diagnostic Test

1. Press the

System

button to enter the system mode.

2. Use the < > arrow keys to move to More, and press

Enter

.

3. Use the < > arrow keys to move to Self-Test, and press

Enter

.

The following failure codes apply:

Table 5-1: Self Test Failure Codes

Power

0

OK

1

5 V Failure

2

15 V Failures

3

Both Failed

Analog

0

OK

1

DAC A (0–1 V Concentration)

2

DAC B (0–1 V Range ID)

3

Both Failed

Preamp

0

OK