2 Installation

Model 7320

2-6

Teledyne Analytical Instruments

as 0–10 % carbon monoxide, then the output would be as shown in
Table 2-1.

Table 2-1: Analog Concentration Output—Example

Percent

Voltage Signal

Current Signal

CO

Output (V dc)

Output (mA dc)

0

0.0

4.0

1

0.1

5.6

2

0.2

7.2

3

0.3

8.8

4

0.4

10.4

5

0.5

12.0

6

0.6

13.6

7

0.7

15.2

8

0.8

16.8

9

0.9

18.4

10

1.0

20.0

To provide an indication of the range, the Range ID analog outputs are

used. They generate a steady preset voltage (or current when using the
current outputs) to represent a particular range. Table 2-2 gives the range ID
output for each analysis range.

Table 2-2: Analog Range ID Output—Example

Range Voltage (V) Current (mA) Application
Range 1 0.25

8 90-100% CO/N

2

Range 2 0.50

12 95-100% CO/N

2

Range 3 0.75

16 0-100% CO/N

2

Range 4 (Cal) 1.00

20 98-100% CO

2

/N

2

2.3.3.2

Alarm Relays

The nine alarm-circuit connector pins connect to the internal alarm relay

contacts. Each set of three pins provides one set of Form C relay contacts.
Each relay has both normally open and normally closed contact connections.
The contact connections are shown in Figure 2-4. They are capable of