Maintenance – Teledyne 3290 - OEM percent oxygen analyzer User Manual

Page 39

background image

Percent Oxygen Analyzer

Maintenance

Teledyne

Analytical

Instruments

29

Maintenance

Aside from normal cleaning and checking for leaks at the gas

connections, the Model 3290 should not require any maintenance
beyond replacement of expended sensors, and perhaps a blown fuse.
Routine maintenance includes occasional recalibration, as described in
Chapter 4, Operation.

5.1 Replacing the Fuse

5.1.1 AC Powered Units

When a fuse blows, check first to determine the cause, then replace

the fuse using the following procedure:

1. Disconnect the AC power and place the power switch located on

the rear panel in the O position. Remove the power cord from the
receptacle.

2. The fuse receptacle is located in the power cord receptacle

assembly in the upper left-hand corner of the rear panel. See
Figure 5-1.

Figure 5-1: AC Fuse Replacement