Operation Model

3290

Teledyne Analytical Instruments

26

4.4.2 LO Alarm

Setting the LO Alarm sets the concentration BELOW which the LO

alarm will activate. To set the LO Alarm:

1. Press the

SET LO ALARM

function button once.

2. Immediately press either the

 or  button to raise or lower the

displayed value, as required, until the display reads the desired
percent concentration.

4.4.3 Sensor Fail Alarm

The SENSOR FAIL alarm is factory preset in the instrument

software at 0.05%. It cannot be changed by the user. It is a low O

2

alarm and triggers when the O

2

level is below this level of 0.05% O

2

.

Teledyne incorporates a low O

2

alarm to indicate sensor failure,

since for a percent O

2

sensor, the most common failure mode is

depletion of the lead anode over time and O

2

exposure, which results in

a zero or near zero signal from the sensor.

Should this alarm trigger, the ALARM Indicator below the SET

function buttons will blink, and the alarm relay contact dedicated to this
function will change state.

4.5 Selecting a Fixed Range or Autoranging

The Model 3290 can operate in fixed high, fixed low, or

autoranging mode. To change modes:

1. Press and then release the

SET HI RANGE

and the

SET LO

RANGE

buttons simultaneously.

2. Immediately (within 5 seconds) press either the

 or  button

until

Auto, Lo, or Hi displays on the LCD, as desired.

After about three seconds, the analyzer resumes monitoring in the

selected range mode.

4.6 Calibration

Preliminary—If not already done: Power up the Analyzer and allow
the LED reading to stabilize. Set the Alarm setpoints and the analysis
ranges to the desired values.