Installation Model

3290

Teledyne Analytical Instruments

20

The specific descriptions for each type of alarm are as follows:

HI Alarm

Configured as high alarm (actuates when
concentration is above threshold). Can be set
anywhere between 1 and 25 %, but must be set
ABOVE the threshold set for the LO Alarm.

LO Alarm

Configured as low alarm (actuates when
concentration is below threshold). Can be set
anywhere from 1 to 25 %, but must be set BELOW
the threshold set for the HI Alarm.

Sensor Fail

Actuates when the output of the oxygen sensor falls
below the pre-programmed level, typically 0.05%..

Figure 3-2: Contact Identification for FAILSAFE Relay Operation

Optional RS-232 Port: The digital signal output is a standard RS-

232 serial communications port used to connect the analyzer to a modem
or other digital device. Only the output mode is implemented in this
instrument. The data is concentration information in serial digital form.

The RS-232 protocol allows some flexibility in implementation in

the choice of values for certain parameters. Table 3-1 lists the RS-232
values required by the Model 3290 implementation.

Note: The RS-232 port is an optional feature and is not present

on all instruments.