background image

Operational Theory

Model 3290

Teledyne Analytical Instruments

12

2.3 Electronics

2.3.1 General

The signal processing uses an Intel

micro-controller with on-board

RAM and ROM to control all signal processing, input/output, and
display functions for the analyzer. System power is supplied from a
universal power supply module designed to be compatible with most
international power sources.

The power supply circuitry is on the Power Supply PCB, which is

mounted vertically, just behind the rear panel of the Control Unit.

The signal processing electronics including the temperature

compensated amplifier, micro-controller, analog to digital and digital to
analog converters are located on the Main PCB, which is mounted
vertically, just behind the front panel of the Control Unit.

2.3.2 Signal Processing

Figure 2-3 is a block diagram of the signal processing electronics

described below.