Percent Oxygen Analyzer

Operational Theory

Teledyne

Analytical

Instruments

13

Figure 2-3: Block Diagram of the Signal Processing Electronics

In the presence of oxygen the cell generates a current. A current to

voltage amplifier (I–E AMPL) converts this current to a voltage.

The second stage amplifier (TEMP COMP) supplies temperature

compensation for the oxygen sensor output. The temperature
compensation amplifier incorporates a thermistor (THERM) that is
physically located in the cell block for the typical oxygen sensor. The
thermistor is a temperature dependent resistance that changes the gain of
the amplifier in proportion to the temperature changes in the block. This
change is inversely proportional to the change in the cell output due to
the temperature changes. As a result there is negligible net change in the
signal due to temperature changes once the sensor comes to equilibrium.
See Specifications in the Appendix.

The output from the temperature compensation amplifier is sent to

an analog to digital converter (ADC), and the resulting digital
concentration signal is sent to the micro-controller.