Trace Oxygen Analyzer

Appendix

Teledyne

Analytical

Instruments

25

Appendix

A.1 Specifications

Ranges: Two user selectable ranges can be set

between 0-10 ppm and 0-1,000 ppm
oxygen. Default ranges are 0-100 ppm and
0-1,000 ppm oxygen

Signal

Output: Voltage: 0–10 VDC, negative ground

Current: 4-20 mA, negative ground

Range ID: 0-10

VDC

Display: Light-emitting diode (LED) display

Alarms: One high alarm relay, adjustable; one low

alarm relay, adjustable; one sensor failure
relay. All alarms failsafe.

System Operating Temp: 0 - 50°C

Accuracy: ±2% of full scale at constant temperature

±5% of full scale over 0-50°C range once
temperature equilibrium is reached (at 3%
and 5% user defined ranges).

Response

Time: 90% in less than 20 seconds at 25°C

System

Power:

AC (100-240 VAC, 47/440 Hz), or DC

Requirements:

(10-36 VDC); user specified

System

Enclosure: Panel Mount: 2.81" H × 6.0" W × 2.87" D

(71.4 mm × 152.4 mm × 72.9 mm).
Face Plate: 3.75" H × 7.0" W (95.3 mm H ×
177.8 mm W). Faceplate rated to NEMA-4

Sensor Type: Zirconium

Oxide

Analysis

Unit: 4.0" H × 6.0" W × 3.25" D

(101.6 mm × 152.4 mm × 82.6 mm)