beautypg.com

Teledyne 3190Z - Zirconium oxide trace oxygen analyzer User Manual

Page 32

background image

Maintenance Model

3190Z

Teledyne Analytical Instruments

24

3. Insert a small flat-blade screwdriver into the slot in the receptacle

wall nearest the fuse and gently pry open the fuse receptacle. The
fuse holder will slide out. The fuse in use is visible in the clip.
To open the spare fuse compartment, push on one end until it
slides out.

4. Remove the bad fuse and replace it with a 5

×20 mm 0.5 A,

250 VAC, IEC, time lag (T) fuse (P/N F1128) for AC units.

5. Replace the fuse holder into its receptacle, pushing in firmly

until it clicks.

5.1.2 DC Powered Units

In units with DC power, the fuse is located on the rear panel above

the ON/OFF switch.

1. Open the fuse holder by unscrewing and removing the cap

marked FUSE.

2. The fuse is located inside the receptacle, not inside the cap. Both

terminals are on the same end of the fuse. Pull straight out
without twisting to remove the old fuse from the receptacle, and
replace it with a 0.5 A, 125 VDC, very quick acting (FF)
Microfuse (P/N F51).

3. Replace the cap by screwing it back into the receptacle.

5.2 Sensor Installation or Replacement

5.2.1 When to Replace a Sensor

The sensors used in this product will typically last five years or
longer in service. Typically, the sensors will fail either high or
low. Inaccurate or erratic readings are a sign that one of the
sensors may have failed. In most cases, we recommend the
entire unit be returned to an authorized Teledyne Service Center
for sensor replacement.