Trace Oxygen Analyzer

Operation

Teledyne

Analytical

Instruments

19

4.3.1 HI Range

Setting the HI Range full-scale value defines the LEAST sensitive

analysis range to be used. To set the HI Range:

1. Press the SET HI RANGE function button once.

2. Immediately (within 5 seconds) press either the

Δ or ∇ button

to raise or lower the displayed value as required, until the
display reads the desired full-scale concentration.

4.3.2 LO Range

Setting the LO Range full-scale value defines the MOST sensitive

range to be used. To set the LO Range:

1. Press the SET LO RANGE function button once.

2. Immediately (within 5 seconds) press either the

Δ or ∇ button

to raise or lower the displayed value as required, until the
display reads the desired full-scale concentration.

4.4 Setting the Alarm Setpoints

The alarm set-points can be adjusted over the full range of the

analyzer (0-1,000 ppm oxygen content). The set-point values are
expressed in ppm only.

Note: The HI Alarm set-point MUST be set at a higher

concentration than the LO Alarm set-point.

4.4.1 HI Alarm

Setting the HI Alarm sets the value ABOVE which the HI Alarm

will activate. To Set the HI Alarm:

1. Press the SET HI ALARM function button once.

2. Within 5 seconds press either the

Δ or ∇ button to raise or lower

the displayed value as required, until the display reads the
desired concentration.

4.4.2 LO Alarm

Setting the LO Alarm sets the value BELOW which the LO alarm

will activate. To set the LO Alarm: