Teledyne 3190Z - Zirconium oxide trace oxygen analyzer User Manual

Page 26

background image

Operation Model

3190Z

Teledyne Analytical Instruments

18

4.2 Using the Function and Data Entry Buttons

When one of the function buttons is being pressed, the analyzer is

in the Setup mode or the Calibration mode.

The four setup function buttons on the analyzer are:

• SET HI ALARM
• SET LO ALARM
• SET HI RANGE
• SET LO RANGE

The Calibration mode button is:

• SPAN

The data entry buttons (

Δ and ∇) adjust the value displayed on the

TRACE OXYGEN meter while one of the function buttons is being held
down.

• Δ : Increments the displayed value upwards.
• ∇ : Increments the displayed value downwards.

Any of the functions can be selected at any time by pressing the

appropriate button.

Each function will be described in the following sections. Although

the operator can use any function at any time, the order chosen in this
manual is appropriate for an initial setup.

4.3 Setting the Analysis Ranges

The two user-definable analysis ranges are both capable of being

adjusted from 0-10 ppm to 0-1,000 ppm oxygen concentration.

The analyzer automatically switches from the LO range to the

HI range when the oxygen concentration reaches the LO range full-
scale value. It will switch back to the LO range when the oxygen
concentration falls below the LO range full-scale value

Note: The HI Range set-point MUST be set at a higher

concentration than the LO Range set-point.