A-9

Percent Oxygen Analyzer

Appendix

Teledyne Analytical Instruments

to toggle between

AUTO and MAN zero settling. Stop when MAN appears,

blinking, on the display.

Zero:

Settling:

Man

To Begin

Press

Enter

to begin the zero calibration. After a few seconds the first

of five zeroing screens appears. The number in the upper left hand corner is
the first-stage zero offset. The microprocessor samples the output at a prede-
termined rate. It calculates the differences between successive samplings and
displays the rate of change as

Slope= a value in parts per million per second

(ppm/s).

#### % Zero

Slope=#### ppm/s

NOTE: It takes several seconds for the true

Slope

Slope

Slope

Slope

Slope

value to display. Wait about

10 seconds. Then, wait until

Slope

Slope

Slope

Slope

Slope

is sufficiently close to zero before

pressing

Enter to finish zeroing .

Generally, you have a good zero when

Slope is less than 0.05 ppm/s

for about 30 seconds. When

Slope is close enough to zero, press

Enter

. In a

few seconds, the screen will update.

Once span settling completes, the information is stored in the

microprocessor, and the instrument automatically returns to the

Analyze

mode.

NOTE: