Teledyne 3000PA - General purpose percent oxygen analyzer User Manual
Page 69

A-9
Percent Oxygen Analyzer
Appendix
Teledyne Analytical Instruments
to toggle between
AUTO and MAN zero settling. Stop when MAN appears,
blinking, on the display.
Zero:
Settling:
Man
Press
Enter
to begin the zero calibration. After a few seconds the first
of five zeroing screens appears. The number in the upper left hand corner is
the first-stage zero offset. The microprocessor samples the output at a prede-
termined rate. It calculates the differences between successive samplings and
displays the rate of change as
Slope= a value in parts per million per second
(ppm/s).
#### % Zero
Slope=#### ppm/s
NOTE: It takes several seconds for the true
Slope
Slope
Slope
Slope
Slope
value to display. Wait about
10 seconds. Then, wait until
Slope
Slope
Slope
Slope
Slope
is sufficiently close to zero before
pressing
Enter to finish zeroing .
Generally, you have a good zero when
Slope is less than 0.05 ppm/s
for about 30 seconds. When
Slope is close enough to zero, press
Enter
. In a
few seconds, the screen will update.
Once span settling completes, the information is stored in the
microprocessor, and the instrument automatically returns to the
Analyze
mode.