Percent Oxygen Analyzer

Maintenance 5

5-3

Teledyne Analytical Instruments

3. Divide the raw output reading by the percent oxygen

concentration of your span gas.

If the quotient is less than the Index value for the cell class you are

using, replace the cell.

Table 5-1: Cell Indices

Cell Class

Index

A-3

1.818

A-5

1.818

B-1

4.545

B-3

3.716

B-5

1.244

B-7

1.515

C-3

2.488

C-5

0.606

5.2.3 Removing the Micro-Fuel Cell

The Micro-Fuel cell is located inside the nylon Probe behind the front

panel. (See Figure 5-1.) To remove an existing cell:

1. Remove power to the instrument by unplugging the power cord

at the power source.

2. Open the front panel door by pressing the release button on the

top right corner of the door all the way in and releasing it.

3. Pull up on the nylon Probe, with a slight rocking motion, to

release it from the Probe Receptacle.

4. Do Not remove the O-rings unless they are worn and no longer

hold the Probe tightly. (If worn, replace them.)

5. When it is free, unscrew the Cap from the nylon Probe. Hold

the Probe vertically to prevent dropping the cell out of the
probe.

6. Remove the Cell from the Probe, and dispose of it in an

environmentally safe manner.