4 Operation

Model 3000PA

4-16

Teledyne Analytical Instruments

4.6.2 Autoranging Analysis

Set your analysis ranges as in 4.6.1, above. Leave

Mode

in

Auto

, or

use the arrow buttons to change back to

Auto

.

When operating in autoranging, if the oxygen concentration in your

sample goes ABOVE your HIGHEST range setting, the analyzer will go
into the special 25 % cal range.

However, if one of your range settings is below 0-25 % and another is

set above 0-25 %, the special 0-25 % Air Cal range will NOT activate as the
oxygen level goes through 25 %. Nevertheless, if the oxygen concentration
in your sample goes ABOVE your HIGHEST range setting, the analyzer
will THEN drop back down into the special 25 % cal range.

Once the oxygen concentration drops back down into your highest

range setting, the analyzer will automatically switch back to that range.

CAUTION: While the analyzer is in the Air Cal range, the oxygen

reading cannot go over 25 %, even if the oxygen
concentration is higher than 25 %.

4.6.3 Fixed Range Analysis

The autoranging mode of the instrument can be overridden, forcing the

analyzer DC outputs to stay in a single predetermined range.

To switch from autoranging to fixed range analysis, enter the range

function by pressing the

Range

button on the front panel.

Use the < > arrow keys to move the blinking over

AUTO

.

Use the

∆∇

arrow keys to switch from

AUTO

to

FX/LO

,

FX/MED,

or

FX/HI

to set the instrument on the desired fixed range (low, medium, or

high).

L—1.00 M—5.00
H—10.00 Mode—FX/LO

or

L—1.00 M—5.00
H—10.00 Mode—FX/MED

or

L—1.00 M—5.00
H—10.00 Mode—FX/HI