Page 41

background image

Percent Oxygen Analyzer

Operation 4

4-9

Teledyne Analytical Instruments

4.3.5 System Self-Diagnostic Test

The Model 3000PA has a built-in self-diagnostic testing routine. Pre-

programmed signals are sent through the power supply, output board and
sensor circuit. The return signal is analyzed, and at the end of the test the
status of each function is displayed on the screen, either as

OK

or as a

number between 1 and 3. (See System Self Diagnostic Test in chapter 5 for
number code.)

The self diagnostics are run automatically by the analyzer whenever the

instrument is turned on, but the test can also be run by the operator at will.
To initiate a self diagnostic test during operation:

Press the

System

button to start the

System

function.

Contrast Auto—Cal
PSWD Logout More

Use the < > arrow keys to blink

More

, then press

Enter

.

Version Self—Test
Cell Output: ### µA

Use the < > arrow keys again to move the blinking to the

Self–Test

function. The screen will follow the running of the diagnostic.

RUNNING DIAGNOSTIC
Testing Preamp — 83

During preamp testing there is a countdown in the lower right corner of

the screen. When the testing is complete, the results are displayed.

Power: OK Analog: OK
Preamp: 3

The module is functioning properly if it is followed by

OK

. A number

indicates a problem in a specific area of the instrument. Refer to Chapter 5
Maintenance and Troubleshooting
for number-code information. The results
screen alternates for a time with:

Press Any Key
To Continue...

Then the analyzer returns to the initial System screen.

4.3.6 Version Screen

Move the < > arrow key to More and press

Enter

. With

Version

blinking, press

Enter

. The screen displays the manufacturer, model, and

software version information.

Contrast Function is

DISABLED

(Refer to Section 1.6)