4 Operation

Model 3000MB

4-16

Teledyne Analytical Instruments

To switch from autoranging to fixed range analysis, enter the range

function by pressing the Range button on the front panel.

Use the < > arrow keys to move the blinking over

AUTO.

Use the

∆∇

arrow keys to switch from

AUTO to FX/LO, FX/MED, or FX/

HI to set the instrument on the desired fixed range (low, medium, or high).

L—1.00 M—5.00
H—10.00 Mode—FX/LO

or

L—1.00 M—5.00
H—10.00 Mode—FX/MED

or

L—1.00 M—5.00
H—10.00 Mode—FX/HI

Press Escape to re-enter the Analyze mode using the fixed range.

NOTE:When performing analysis on a fixed range, if the oxygen

concentration rises above the upper limit (or default value) as
established by the operator for that particular range, the
output saturates at 1 V dc. However, the digital readout and
the RS-232 output continue to read the true value of the oxy-
gen concentration regardless of the analog output range.

4.7

The Analyze Function

Normally, all of the functions automatically switch back to the Analyze

function when they have completed their assigned operations. Pressing the
Escape

button in many cases also switches the analyzer back to the Ana-

lyze

function. Alternatively, you can press the Analyze button at any time

to return to analyzing your sample.

4.8

Signal Output

The standard Model 3000MB Percent Oxygen Analyzer is equipped

with two 0–1 V dc analog output terminals accessible on the back panel (one
concentration and one range ID). The –MA option also has two isolated 4–
20 mA dc current outputs (one concentration and one range ID).

See Rear Panel in Chapter 3, Installation, for illustration.