Percent Oxygen Analyzer

Operation 4

4-11

Teledyne Analytical Instruments

Press Enter to move to the next screen.

Span Val: 20.90
Span Mod #

Use the

∆∆∆∆∆∇

arrow keys to enter the oxygen-concentration mode. Use

the < > arrow keys to blink the digit you are going to modify. Use the

∆∆∆∆∆∇

arrow keys again to change the value of the selected digit. When you have
finished typing in the concentration of the span gas you are using (20.90 if
you are using air), press Enter to begin the Span calibration.

#### % Span
Slope=#### ppm/s

The beginning span value is shown in the upper left corner of the

display. As the span reading settles, the screen displays and updates informa-
tion on

Slope. Spanning automatically ends when the slope is less than 1/50

of the displayed value of the oxygen concentration (in ppm) for three min-
utes. Then the instrument automatically returns to the analyze mode.

4.4.1.2

Manual Mode Spanning

Press Span to start the Span function. The screen that appears allows

you to select whether the span calibration is to be performed automatically or
manually.

Span: Settling:MAN
For Next

Use the

∆∇

keys to toggle between

AUTO and MAN span settling.

Stop when

MAN appears, blinking, on the display. Press Enter to move to

the next screen.

Span Val: 20.90
Span Mod #

Press

(

) to permit modification (Mod #) of span value.

Use the arrow keys to enter the oxygen concentration of the span gas

you are using (20.90 if you are using air). The < > arrows chose the digit,
and the

∆∇

arrows choose the value of the digit.

Press Enter to enter the span value into the system and begin the span

calibration.

Once the span has begun, the microprocessor samples the output at a

predetermined rate. It calculates the difference between successive samplings
and displays this difference as Slope on the screen. It takes several seconds