beautypg.com

Teledyne 3000MB - Paramagnetic oxygen analyzer User Manual

Page 33

background image

Percent Oxygen Analyzer

Installation 3

3-13

Teledyne Analytical Instruments

Figure 3-10: Gas Connector Panel

SAMPLE IN: In the standard model, gas connections are made at the

SAMPLE IN

and

EXHAUST OUT

connections. Calibration gases must be

Tee'd into the Sample inlet with appropriate valves.

The gas pressure in should be reasonably regulated. Pressures between

3 and 40 psig are acceptable as long as the pressure, once established, will
keep the front panel flowmeter reading in an acceptable range (0.1 to 2.4
SLPM). Exact figures will depend on your process.

If greater flow is required for improved response time, install a bypass

in the sampling system upstream of the analyzer input.

Note: If you have the –V option, The above numbers apply instead to

the vacuum at the EXHAUST OUT connector, described be-
low, with minus signs before the pressure readings.

ZERO IN and SPAN IN: These are additional ports, included on

models with the -C option, for inputting span gas and zero gas. There are
electrically operated valves inside for automatic switching between sample
and calibration gases. These valves are completely under control of the
3000PB Electronics. They can be externally controlled only indirectly
through the Remote Cal Inputs, described below.

Behind Panel