beautypg.com

Teledyne 3000MB - Paramagnetic oxygen analyzer User Manual

Page 46

background image

4 Operation

Model 3000MB

4-12

Teledyne Analytical Instruments

for the first Slope value to display. Slope indicates rate of change of the
Span

reading. It is a sensitive indicator of stability.

#### % Span
Slope=#### ppm/s

When the Span value displayed on the screen is sufficiently stable,

press Enter. (Generally, when the Span reading changes by 1 % or less of
the full scale of the range being calibrated for a period of five minutes it is
sufficiently stable.) Once Enter is pressed, the Span reading changes to the
correct value. The instrument then automatically enters the Analyze func-
tion.

4.5

The Alarms Function

The Model 3000MB is equipped with 2 fully adjustable concentration

alarms and a system failure alarm. Each alarm has a relay with a set of form
“C" contacts rated for 3 amperes resistive load at 250 V ac. See Figure in
Chapter 3, Installation and/or the Interconnection Diagram included at the
back of this manual for relay terminal connections.

The system failure alarm has a fixed configuration described in chapter

3 Installation.

The concentration alarms can be configured from the front panel as

either high or low alarms by the operator. The alarm modes can be set as
latching or non-latching, and either failsafe or non-failsafe, or, they can be
defeated altogether. The setpoints for the alarms are also established using
this function.

Decide how your alarms should be configured. The choice will depend

upon your process. Consider the following four points:

1. Which if any of the alarms are to be high alarms and which if any

are to be low alarms?
Setting an alarm as HIGH triggers the alarm when the oxygen
concentration rises above the setpoint. Setting an alarm as LOW
triggers the alarm when the oxygen concentration falls below the
setpoint.
Decide whether you want the alarms to be set as:

Both high (high and high-high) alarms, or

One high and one low alarm, or

Both low (low and low-low) alarms.

2. Are either or both of the alarms to be configured as failsafe?