beautypg.com

Range – Teledyne 3000MA - Paramagnetic oxygen analyzer User Manual

Page 44

background image

4 Operation

Model 3000MA

4-14

Teledyne Analytical Instruments

To define the setpoint, use the < > arrow keys to move the
blinking over to AL–1 ####. Then use the

∆∇

arrow keys to

change the number. Holding down the key speeds up the
incrementing or decrementing. (Remember, the setpoint units are
percent-of-oxygen.)

To set the other parameters use the < > arrow keys to move the
blinking over to the desired parameter. Then use the

∆∇

arrow

keys to change the parameter.

Once the parameters for alarm 1 have been set, press Alarms
again, and repeat this procedure for alarm 2 (AL–2).

To reset a latched alarm, go to

Dft– and then press either

two

times or

two times. (Toggle it to

Y and then back to N.)

–OR –

Go to

Ltch– and then press either

two times or

two times.

(Toggle it to

N and back to Y.)

4.6

The

Range

Function

The Range function allows the operator to program up to three concen-

tration ranges to correlate with the DC analog outputs. If no custom ranges
are defined by the user, the instrument defaults to:

Low = 0–1.00 %
Med = 0–5.00 %
High = 0–10.00 %.

The Model 3000MA is set at the factory to default to autoranging. In

this mode, the microprocessor automatically responds to concentration
changes by switching ranges for optimum readout sensitivity. If the current
range limits are exceeded, the instrument will automatically shift to the next
higher range. If the concentration falls slightly below full scale of the next
lower range, the instrument will switch to that range. A corresponding shift
in the DC percent-of-range output, and in the range ID outputs, will be
noticed.

The autoranging feature can be overridden so that analog output stays

on a fixed range regardless of the oxygen concentration detected. If the
concentration exceeds the upper limit of the range, the DC output will
saturate at 1 V dc.

However, the digital readout and the RS-232 output of the concentra-

tion are unaffected by the fixed range. They continue to read accurately with
full precision. See Front Panel description in Chapter 1.