1-6

1 Introduction

Model 3000MA

Teledyne Analytical Instruments

Alarm Connections

2 concentration alarms and 1 system
alarm.

RS-232 Port

Serial digital concentration signal output
and control input.

Remote Probe

Used in the 3000MA for controlling
external solenoid valves only.

Remote Span/Zero

Digital inputs allow external control of
analyzer calibration. (See Note, below.)

Calibration Contact

To notify external equipment that
instrument is being calibrated and
readings are not monitoring sample.

Range ID Contacts

Four separate, dedicated, range relay
contacts. Low, Medium, High, Cal.

Network

For future expansion. Not implemented
at this printing.

Optional:

Calibration Gas Ports

Separate fittings for zero, span and
sample gas input, and internal valves
for automatically switching the gases.

Current Signal Output Additional isolated 4-20 mA dc plus

4-20 mA dc range ID.

Note: If you require highly accurate Auto-Cal timing, use external

Auto-Cal control where possible. The internal clock in the
Model 3000MA is accurate to 2-3 %. Accordingly, internally
scheduled calibrations can vary 2-3 % per day.