Charging Time
The Battery Charger will completely charge a battery in
less than 3 hours. (If the optional

Dual Bay Battery

Charger is used with two batteries inserted, they will
charge sequentially. The status lamp will be green for a
charged battery and yellow for a battery waiting to be
charged. The second battery will automatically
commence charging once the first battery is fully
charged.)

Note: The 2240

Battery Charger monitors the

battery temperature and will not commence charging
a battery that is outside the temperature range of 10
to 45 °C. The yellow lamp will illuminate and remain
illuminated as long as the battery temperature is
outside this range.

An error indication (flashing red light) will be given if
the battery terminals are shorted or if the battery fails to
properly take a charge.

In Case of a Problem
If the Battery Charger does not function as expected,
check the following:
• Make sure the wall adapter is properly connected.
• Make sure there are no foreign objects lodged in the

battery bay pockets preventing the battery from
properly seating.

• Make sure the batteries are installed so that they

properly mate with the connector in the bottom of the
pocket.


Page 26 of 26