Teledyne 2240 – Portable Handheld Hydrogen Leak Detector, 2nd generation User Manual

Page 10

background image

Page 19 of 26

M2240 R0

both 0-5 V and 4-20 mA, and the percent of hydrogen
being detected.

Note: before proceeding with this function, first
select the appropriate Analog Output Configuration
Mode.

• Press the Enter button while “Analog Full Scale” is

displayed.

• Specify the hydrogen concentration represented by both

the maximum voltage from the 0-5 V Analog Output and
the maximum current from the 4-20 mA Analog Output
(Default, 2% hydrogen). The minimum voltage and
current value is factory set to represent 0% hydrogen,
and is not configurable.
• For example, using the second “0-5 Volt Config”

format of “0.5-4.5 Volt Mode”, the operator chooses to
designate the maximum voltage output, 4.5 V, to
indicate 4% hydrogen in air. This results in 4.5 V =
4% hydrogen, 0.5 V= 0% hydrogen, and 2.5 V = 2%
hydrogen.

• For example, using the 4-20 mA Analog Output, the

operator chooses 20 mA to indicate 10% hydrogen in
air. This results in 20 mA = 10% hydrogen, 4 mA =
0% hydrogen, and 12 mA = 5% hydrogen.

• Use the Scroll buttons to increase or decrease the percent

value. Use the Mode button to change the digit affected
by the scroll buttons.

• Press the Enter key to set the displayed number into

memory and return to the Configuration Mode.


Backlight – Controls how the LCD backlight operates.

• Press the Enter button while "Backlight" is displayed.
• Select one of three options:

• Always Off
• Always On

Page 20 of 26

M2240 R0

• On With Change - Under this option, the backlight

illuminates for 15 seconds when any button is
pushed, or on any Alarm. It also illuminates
continuously in configuration mode.

• Press the Enter button to select the option.
• Press the Mode button to exit without making any

change.


Rev X.xx – The controller firmware revision level is

displayed.