3 Installation

Model 1220

3-4

Teledyne Analytical Instruments

Figure 3-2 Maximum Loop Resistance for Sensor Connection

The terminals marked "Ext. Audible Alarm" provide an external signal

that is activated whenever the audible alarm is activated. By setting jumpers
on the Control Unit PCB (see section 3.3), the external signal can be either
(1) a contact closure, or (2) a powerline level signal.

Reactive loads connected to the relay contacts must be equipped with

appropriate transient suppression networks to prevent spurious triggering of
sensitive alarm circuits.

Analog output from each channel module should be connected to

monitoring equipment as shown in Figure 3-3. Note that the monitoring
equipment inputs should be floating (not grounded). The cable shield should
be connected to ground only at one end of the cable.

Figure 3-3 Analog Output Connection