Mechanical connections – Perma Pure Micro-GASS UG-1212 User Manual

Page 12

background image

Figure 3 – Electrical Connections

Mechanical Connections

All connections to the MicroGASS system are intended to be made with Teflon or

other suitable plastic tubing

Wet Sample Gas Inlet -

Compression fitting for ¼″ OD Tubing. To be connected to the

sample gas source via a heat traced line to prevent condensation of the sample gas. Once
the heated sample line is installed, the heated line inlet fitting can be shrunk into place
permanently.


Heat shrink
sealing fitting

Figure 4 – Heated Line Connection

Calibration Gas Inlet -

Compression fitting for ¼″ OD Tubing. To be connected to the

sample calibration gas source. Do not exceed 30 psi. Port must be capped when not in use

Dry Sample Gas Outlet -

Compression fitting for ¼″ OD Tubing. To be connected to

analyzer’s inlet port