Control Company 4430 KANGAROO THERMOMETER User Manual
Control company, Traceable, Kangaroo™ thermometer instructions

Traceable
®
is a registered trademark of Control Company ©2008 Control Company. 92-4430-00 Rev. 1 052108
SPECIFICATIONS 
Range: 
-58 to 572 °F / -50 to 300 °C
Resolution: 0.1° (-19.9 to 199.9) 1° (otherwise) 
Features: 
High/Low Temperature Alarm Minimum/Maximum
Memory 
 
PROBE CABLE COMPARTMENT 
Open the cable compartment located at the top of the back of the unit 
to access the probe cable. The bottom of the compartment is hinged 
and will drop down. Open the bottom portion first. The top portion of 
the compartment is designed to be pulled straight out approximately 
1/4 inch allowing the probe cable to be un-wound/wound around the 
cable spool. The top portion is not designed to be pulled out 
completely. 
 
OPERATION 
Remove the plastic tip protector from the tip of the stainless steel 
probe. Press the ON/OFF button to turn the unit on. Press the °F/°C 
button to switch between Fahrenheit and Celsius temperature. Insert 
the probe into the material to be measured and read the display. 
Note:
The probe handle and cable are not waterproof. Do not immerse probe
handle or cable in liquid. Immersing the probe handle or cable in liquid can 
cause permanent damage to the thermometer. 
To turn the unit off, press the ON/OFF button. To conserve battery 
life, always turn the unit off when it is not in use. 
 
MINIMUM/MAXIMUM MEMORIES 
While taking measurements, press the MAX/MIN button once. The 
minimum temperature tested will be displayed. Press the MAX/MIN 
button again and maximum temperature tested will be displayed. 
The display will return to current temperature automatically after 
three seconds. 
The minimum/maximum memories are cleared when the 
thermometer is turned off. 
 
TEMPERATURE ALARM 
While taking measurements, press the LO AL or HI AL button to set 
an alarm at a desired minimum (LO AL) or maximum (HI AL) 
temperature limit. 
Press and hold the button to rapidly advance the display. 
Once the desired alarm temperature appears on the display, wait 3 
seconds, the temperature alarm point will be set and the display will 
return to the current temperature reading. “LOW” and/or “HI” will 
show on the display indicating the alarm has been set. 
The alarm will sound 4 beeps per minute when the measured 
temperature equals or is greater than the maximum or less than the 
minimum set values. To silence/stop the alarm, press any button 
while the unit is alarming. 
 
To clear the alarm setting, press the °F/°C or ON/OFF button. 
 
 
HOLD/TEST FUNCTION
While taking measurements, press the H/T (HOLD/TEST) button to 
“freeze” the reading for recording purposes (“HOLD” will flash on the 
display). Press the H/T button again to resume testing. 
 
Note:
While the HOLD function is being used, the thermometer will not
measure/record any temperature values for the Minimum/Maximum 
Memories, or High/Low Alarms. 
 
BENCH STAND 
The unit is supplied with a bench stand that is part of the back of the 
unit. To use the bench stand, locate the small rectangular opening at 
the bottom of the unit. Place your fingernail into the opening and flip 
the stand out. To close the stand, simply snap it shut. 
 
ALL OPERATIONAL DIFFICULTIES 
If this thermometer does not function properly for any reason, replace 
the battery with a new high quality battery (see “Battery Replacement” 
section). Low battery power can cause any number of “apparent” 
operational difficulties. Replacing the battery with a new fresh battery 
will solve most difficulties. 
 
BATTERY REPLACEMENT 
Erratic readings, a faint display, no display, or appearing on the display 
are all indications that the battery must be replaced. Flip open the 
bench stand and slide the battery cover off in the direction of the arrow. 
Remove the exhausted battery and replace it with a new 1.5 volt silver-
oxide 357 size battery. Make certain positive (+) side is visible. 
Replace battery cover. Close the bench stand. 
 
 
 
 
 
 
 
 
 
 
WARRANTY, SERVICE, OR RECALIBRATION 
For warranty, service, or recalibration, contact:
CONTROL COMPANY
4455 Rex Road
Friendswood, Texas 77546 USA
Ph. 281-482-1714 Fax 281-482-9448
E-mail [email protected]
www.control3.com
Control Company is ISO 9001 Quality-
Certified by DNV and ISO 17025 accredited
as a Calibration Laboratory by A2LA.
TRACEABLE
®
KANGAROO™ THERMOMETER INSTRUCTIONS