Nova-Tech Digi–Sense 12-Channel Benchtop Scanning Thermometer User Manual
Page 56

55
Front panel displays and controls continued
How to calibrate to water
How to calibrate to water
Ice Point
Calibrate CALPT1 to the freezing point of water (0
°
C or 32
°
F) by using
crushed ice made of distilled water. Immerse the probe in a flask
(insulated) filled with crushed ice and topped off with distilled water.
Add crushed ice to keep the flask full of crushed ice and water during
calibration.
Boiling Point
Calibrate CALPT2 to the boiling point of water (100
°
C or 212
°
F) by using
distilled water and correcting for altitude and barometric pressure. As a
rule of thumb the boiling point of water decreases 1
°
C or 1.8
°
F for every
1000 feet above sea level. Correction for barometric pressure is usually
required only in the case of severe high or low pressure weather condi-
tions.
Note: This method is less accurate than using a dry block calibrator or a
thermocouple calibration furnace. Boiling point calibration is only recom-
mended when this equipment is not available and two point calibration is
required.
BA 50-59
10/12/01, 11:07 AM
55
- B4000-16 (28 pages)
- B4000-M (23 pages)
- Foxx Life Sciences (2 pages)
- CON 700 (24 pages)
- pH-CON 700 (32 pages)
- 2700 Series Benchtop Meters (60 pages)
- Oakton Testr 35 (6 pages)
- 4066, 4067, 4068, 4069, 4161, 4162, 4173, 4174, 4176, 4177 Control Company (4 pages)
- 4065, 4175, 4270, 4271, 4274, 4565, 4570, 4574 Control Company (4 pages)
- 4170 Control Company (4 pages)
- WD-35653-15 (3 pages)
- WD-35653-13 (1 page)
- pH-CON 300 (26 pages)
- pH-CON 10 (29 pages)
- CON 6+ (40 pages)
- 4070 Control Company (1 page)
- WD-00653-00 Zero Oxygen (1 page)
- pH-Dissolved Oxygen Meter (35 pages)
- DO 700 (36 pages)
- DO 6+ (48 pages)
- Foxx Autofil Bottle Top Filtration (2 pages)
- Foxx Autofil Centrifuge Funnel and Stand (2 pages)
- Foxx Autofil Laboratory Storage Bottles (2 pages)
- Foxx Autofil Pedestal and Cradle Ring (4 pages)
- RITEFLOW FLOWMETERS AND KITS (16 pages)
- Roto-Flo Flow Indicators 19937 Series (1 page)
- Leveling Base Kit (each) H40404-1000 (1 page)
- Taylor M-3000 TTi3000 Colorimeter Series (28 pages)
- Bel-Art Magnetic Stirring Bars Selection & Use Guide (4 pages)
- VISPOL-Series Polarimeters (2 pages)
- UniPol L-Series Polarimeters (4 pages)
- iPR-Series (8 pages)
- Atago AP-300 Series Polarimeter (4 pages)
- Foxx Heavy Duty Vacuum Bottle (2 pages)
- pH 700 Ion 700 (18 pages)
- pH Tutor Benchtop Meter (2 pages)
- pH 4.01 Buffer (1 page)
- pH 5+ (28 pages)
- pH 11 (76 pages)
- pH 300 (24 pages)
- 4080 TRACEABLE HUMIDITY_TEMPERATURE DEW POINT METER (2 pages)
- 4088 TRACEABLE HYGROMETER_THERMOMETER (1 page)
- 3651 TRACEABLE ANEMOMETER_HYGROMETER_THERMOMETER_DEW POINT PEN (2 pages)
- 3652 TRACEABLE ANEMOMETER_HYGROMETER_THERMOMETER_DEW POINT BAROMETER PEN (3 pages)
- Sibata 6168 Filter Holders (1 page)