How to do a field calibration
Factory calibration is stored in protected, non-volatile memory. Field
calibration is usually used to reduce probe errors over a specified range.
All field calibration values may be cleared (returning the instrument to
factory calibration) by pushing and holding CLEAR during the power up
display test.
An individual channel's field calibration may be cleared. When CALPT1
is displayed, press and hold CLEAR until CALPT1 CLR is displayed. This
also clears CALPT2 for this same channel. The CAL icon will also go off
for this channel.
There are no restrictions on what values may be entered for field
calibration.
Two calibration points may be entered for each channel. After CALPT1 is
entered the prompt for CALPT2 appears.
If only CALPT1 is entered the offset is affected.
If both CALPT1 and CALPT2 are entered, both slope and offset are
affected.
(See calibrating to the freezing/boiling points of water on page 53, for an
example.)
Front panel displays and controls continued
How to do a field calibration
NUM
°
F
CAL
5
(continued on next page)
BA 50-59
10/12/01, 11:07 AM
50
- B4000-16 (28 pages)
- B4000-M (23 pages)
- Foxx Life Sciences (2 pages)
- CON 700 (24 pages)
- pH-CON 700 (32 pages)
- 2700 Series Benchtop Meters (60 pages)
- Oakton Testr 35 (6 pages)
- 4066, 4067, 4068, 4069, 4161, 4162, 4173, 4174, 4176, 4177 Control Company (4 pages)
- 4065, 4175, 4270, 4271, 4274, 4565, 4570, 4574 Control Company (4 pages)
- 4170 Control Company (4 pages)
- WD-35653-15 (3 pages)
- WD-35653-13 (1 page)
- pH-CON 300 (26 pages)
- pH-CON 10 (29 pages)
- CON 6+ (40 pages)
- 4070 Control Company (1 page)
- WD-00653-00 Zero Oxygen (1 page)
- pH-Dissolved Oxygen Meter (35 pages)
- DO 700 (36 pages)
- DO 6+ (48 pages)
- Foxx Autofil Bottle Top Filtration (2 pages)
- Foxx Autofil Centrifuge Funnel and Stand (2 pages)
- Foxx Autofil Laboratory Storage Bottles (2 pages)
- Foxx Autofil Pedestal and Cradle Ring (4 pages)
- RITEFLOW FLOWMETERS AND KITS (16 pages)
- Roto-Flo Flow Indicators 19937 Series (1 page)
- Leveling Base Kit (each) H40404-1000 (1 page)
- Taylor M-3000 TTi3000 Colorimeter Series (28 pages)
- Bel-Art Magnetic Stirring Bars Selection & Use Guide (4 pages)
- VISPOL-Series Polarimeters (2 pages)
- UniPol L-Series Polarimeters (4 pages)
- iPR-Series (8 pages)
- Atago AP-300 Series Polarimeter (4 pages)
- Foxx Heavy Duty Vacuum Bottle (2 pages)
- pH 700 Ion 700 (18 pages)
- pH Tutor Benchtop Meter (2 pages)
- pH 4.01 Buffer (1 page)
- pH 5+ (28 pages)
- pH 11 (76 pages)
- pH 300 (24 pages)
- 4080 TRACEABLE HUMIDITY_TEMPERATURE DEW POINT METER (2 pages)
- 4088 TRACEABLE HYGROMETER_THERMOMETER (1 page)
- 3651 TRACEABLE ANEMOMETER_HYGROMETER_THERMOMETER_DEW POINT PEN (2 pages)
- 3652 TRACEABLE ANEMOMETER_HYGROMETER_THERMOMETER_DEW POINT BAROMETER PEN (3 pages)
- Sibata 6168 Filter Holders (1 page)