Error messages – Nova-Tech DO 6+ User Manual
Page 38

Instruction Manual
DO 6+
- 32 -
8. ERROR
MESSAGES
LCD Display
Indicates Cause
Solution
Low battery
Need new batteries or
battery connection is
bad.
Clean battery contacts.
Replace batteries,
noting polarity.
“Err 1” in %
Saturation
Mode
% Saturation
calibration
error
Calibration is attempted
when the factory
calibrated absolute
value is within
10.1% to 49.9%
Check the value of the
calibration solution. If
zero calibration is done,
make sure the limit of
10% is not exceeded
Recondition your probe
“Err. 1” in mg/L
(ppm)
Concentration
Mode
Concentration
calibration
error
Calibration is attempted
when the factory
calibrated absolute
value is below 2.00
Verify the solution is
above 2.00
Verify the temperature
and salinity settings
Recondition your probe
“UR”/”OR” with
blinking “A”
annunciator in
Temperature
Mode
ATC probe
error,
Under Range,
Over Range
ATC probe is
disconnected or broken
with the ATC feature
activated.
Temperature is out of
range
Connect the ATC plug
to the meter. Verify the
temperature accuracy.
Ensure probe is not
broken or punctured.
“----“with
blinking “A”
annunciator in
% Saturation
and
Concentration
Mode
ATC probe
error
ATC probe is
disconnected or broken
with the ATC feature
activated.
Temperature is out of
range
Connect the ATC plug
to the meter. Verify the
temperature accuracy.
Ensure probe is not
broken or punctured.