Nova-Tech DO 6+ User Manual
Page 18

Instruction Manual
DO 6+
- 12 -
3.3.2 0%
Calibration
1) Rinse the probe well with DI water or
rinse solution.
2) Press
MODE to select % saturation.
3) Dip the probe into zero oxygen
solution. Wait for the reading to
stabilize.
4) Press
CAL. The CAL indicator and
intended percentage calibration point
(0%) display briefly, before CAL
flashes. The % value based on the
factory default calibration is shown.
5) Press
HOLD/ENTER to confirm the
calibration. The meter displays “CO”,
automatically calibrates to 0.0%
saturation, then returns to
measurement mode.
A value with respect to the
factory default calibration
will be displayed
%
CAL
A
%
HOLD
ENTER
CAL
CAL
%
Calibration point will be
displayed for 1.5 seconds
%
“CO” will displayed for
1.5 seconds
A
See also other documents in the category Nova-Tech Equipment:
- B4000-16 (28 pages)
- B4000-M (23 pages)
- Foxx Life Sciences (2 pages)
- CON 700 (24 pages)
- pH-CON 700 (32 pages)
- 2700 Series Benchtop Meters (60 pages)
- Oakton Testr 35 (6 pages)
- 4066, 4067, 4068, 4069, 4161, 4162, 4173, 4174, 4176, 4177 Control Company (4 pages)
- 4065, 4175, 4270, 4271, 4274, 4565, 4570, 4574 Control Company (4 pages)
- 4170 Control Company (4 pages)
- WD-35653-15 (3 pages)
- WD-35653-13 (1 page)
- pH-CON 300 (26 pages)
- pH-CON 10 (29 pages)
- CON 6+ (40 pages)
- 4070 Control Company (1 page)
- WD-00653-00 Zero Oxygen (1 page)
- pH-Dissolved Oxygen Meter (35 pages)
- DO 700 (36 pages)
- Foxx Autofil Bottle Top Filtration (2 pages)
- Foxx Autofil Centrifuge Funnel and Stand (2 pages)
- Foxx Autofil Laboratory Storage Bottles (2 pages)
- Foxx Autofil Pedestal and Cradle Ring (4 pages)
- RITEFLOW FLOWMETERS AND KITS (16 pages)
- Roto-Flo Flow Indicators 19937 Series (1 page)
- Leveling Base Kit (each) H40404-1000 (1 page)
- Taylor M-3000 TTi3000 Colorimeter Series (28 pages)
- Bel-Art Magnetic Stirring Bars Selection & Use Guide (4 pages)
- VISPOL-Series Polarimeters (2 pages)
- UniPol L-Series Polarimeters (4 pages)
- iPR-Series (8 pages)
- Atago AP-300 Series Polarimeter (4 pages)
- Foxx Heavy Duty Vacuum Bottle (2 pages)
- pH 700 Ion 700 (18 pages)
- pH Tutor Benchtop Meter (2 pages)
- pH 4.01 Buffer (1 page)
- pH 5+ (28 pages)
- pH 11 (76 pages)
- pH 300 (24 pages)
- 4080 TRACEABLE HUMIDITY_TEMPERATURE DEW POINT METER (2 pages)
- 4088 TRACEABLE HYGROMETER_THERMOMETER (1 page)
- 3651 TRACEABLE ANEMOMETER_HYGROMETER_THERMOMETER_DEW POINT PEN (2 pages)
- 3652 TRACEABLE ANEMOMETER_HYGROMETER_THERMOMETER_DEW POINT BAROMETER PEN (3 pages)
- Sibata 6168 Filter Holders (1 page)