2 single or multi-point calibration, Single or multi-point calibration – Nova-Tech CON 700 User Manual
Page 9
5
The factory default is automatic conductivity calibration. See Section
6.10
to change this setting. There is no automatic calibration value
available for conductivity range 1.
TABLE 1
Automatic Calibration Values
Normalization Temperature
Range
#
Conductivity
Range
25 ºC
20 ºC
r 1
0.00 – 20.00 μS None
None
r 2
20.1 – 200.0 μS 84
μS 76
μS
r 3
200.1 – 2000 μS 1413
μS 1278
μS
r 4
2.01 – 20.00 mS
12.88 mS
11.67 mS
r 5
20.1 – 200.0 mS
111.8 mS
102.1 mS
Range
#
TDS Range
(using 0.5 TDS factor)
Automatic Calibration
Values
r 1
0 – 10.00 ppm
none
r 2
10.1 – 100.0 ppm
none
r 3
101 – 1000 ppm
none
r 4
1.01 – 10.00 ppt
none
r 5
10.1 – 100 ppt
none
4.2 Single or Multi-Point Calibration
Use Single-Point Calibration to apply a single calibration value across
all ranges.
Use Multi-Point Calibration for individual calibration in each range. This
will restrict an individual calibration so that it is applied to one range
only. When using multi-point calibration, perform a calibration in each
range that you expect to use for best results.