7.
CALCULATING TDS CONVERSION FACTOR
You can calibrate TDS using the value of the calibration standard solution
at a standard temperature such as 25 ºC. To determine the conductivity-
to-TDS conversion factor use the following formula:
Factor = Actual TDS ÷ Actual Conductivity @ 25 ºC
Actual TDS: Value from the solution bottle label or as a
standard made using high purity water and precisely weighed
salts.
ppm = milligram of salt(s) per liter of water = mg/L
ppt = gram of salt(s) per liter of water = g/L
Actual Conductivity: Value measured using a properly
calibrated Conductivity/Temperature meter.
Both the Actual TDS and the Actual Conductivity values must be in the
same magnitude of units. For example, if the TDS value is in ppm the
conductivity value must be in µS; if the TDS value is in ppt the conductivity
value must be in mS.
Check your factor by multiplying the conductivity reading by the factor in
the above formula. The result should be in TDS value.
When the TDS factor is set to 1.0, Conductivity = TDS.
- B4000-16 (28 pages)
- B4000-M (23 pages)
- Foxx Life Sciences (2 pages)
- pH-CON 700 (32 pages)
- 2700 Series Benchtop Meters (60 pages)
- Oakton Testr 35 (6 pages)
- 4066, 4067, 4068, 4069, 4161, 4162, 4173, 4174, 4176, 4177 Control Company (4 pages)
- 4065, 4175, 4270, 4271, 4274, 4565, 4570, 4574 Control Company (4 pages)
- 4170 Control Company (4 pages)
- WD-35653-15 (3 pages)
- WD-35653-13 (1 page)
- pH-CON 300 (26 pages)
- pH-CON 10 (29 pages)
- CON 6+ (40 pages)
- 4070 Control Company (1 page)
- WD-00653-00 Zero Oxygen (1 page)
- pH-Dissolved Oxygen Meter (35 pages)
- DO 700 (36 pages)
- DO 6+ (48 pages)
- Foxx Autofil Bottle Top Filtration (2 pages)
- Foxx Autofil Centrifuge Funnel and Stand (2 pages)
- Foxx Autofil Laboratory Storage Bottles (2 pages)
- Foxx Autofil Pedestal and Cradle Ring (4 pages)
- RITEFLOW FLOWMETERS AND KITS (16 pages)
- Roto-Flo Flow Indicators 19937 Series (1 page)
- Leveling Base Kit (each) H40404-1000 (1 page)
- Taylor M-3000 TTi3000 Colorimeter Series (28 pages)
- Bel-Art Magnetic Stirring Bars Selection & Use Guide (4 pages)
- VISPOL-Series Polarimeters (2 pages)
- UniPol L-Series Polarimeters (4 pages)
- iPR-Series (8 pages)
- Atago AP-300 Series Polarimeter (4 pages)
- Foxx Heavy Duty Vacuum Bottle (2 pages)
- pH 700 Ion 700 (18 pages)
- pH Tutor Benchtop Meter (2 pages)
- pH 4.01 Buffer (1 page)
- pH 5+ (28 pages)
- pH 11 (76 pages)
- pH 300 (24 pages)
- 4080 TRACEABLE HUMIDITY_TEMPERATURE DEW POINT METER (2 pages)
- 4088 TRACEABLE HYGROMETER_THERMOMETER (1 page)
- 3651 TRACEABLE ANEMOMETER_HYGROMETER_THERMOMETER_DEW POINT PEN (2 pages)
- 3652 TRACEABLE ANEMOMETER_HYGROMETER_THERMOMETER_DEW POINT BAROMETER PEN (3 pages)
- Sibata 6168 Filter Holders (1 page)