beautypg.com

5 manual conductivity & tds calibration procedure, 6 temperature calibration, Measurement – Nova-Tech CON 700 User Manual

Page 11: 1 taking measurements, Manual conductivity & tds calibration procedure, Temperature calibration, Taking measurements

background image

4.5 Manual Conductivity & TDS Calibration Procedure

1) Press as needed to select conductivity (μS or mS) or TDS

(ppm or ppt) calibration.

2) Dip the electrode into the calibration standard and press

.

Provide stirring for best results.

3) The primary display will show the current reading, while the

secondary display will be the factory default value. Adjust the

temperature reading using

or

. Press

to accept. The

primary reading will flash briefly before returning to measurement
mode upon successful calibration.


4.6 Temperature Calibration
The thermistor sensor used for automatic temperature compensation
and measurement is accurate and stable, so frequent calibration isn’t
required. Temperature calibration is recommended upon probe
replacement, whenever the temperature reading is suspect, or if
matching against a certified thermometer is desired.

1) Place the probe into a solution with a known accurate temperature

such as a constant temperature bath.

2) Press

followed by

. The primary display shows the

measured temperature while the secondary display shows the
factory default temperature.

4) Adjust the temperature using

or

. Press

to accept or

to cancel. The meter allows an adjustable maximum value of ± 5

º

C

(or ± 0.9

º

F) from the factory default temperature.

5.

MEASUREMENT

5.1 Taking Measurements
1)

Rinse the electrode with de-ionized or distilled water before use
to remove any impurities. Gently shake excess water droplets.

7