Troubleshooting guide – Nova-Tech CON 700 User Manual
Page 20

10.
TROUBLESHOOTING GUIDE
PROBLEM CAUSE
SOLUTION
No display
Main power not switched
on. AC Adapter socket not
inserted properly.
a) Switch on the power supply.
b) Re-insert AC Adapter.
“Ur” (Under
range)
“Or” (Over
range)
in primary or
secondary
display
Measured value is out
of range.
Electrodes not connected.
Electrode clogged, dirty
or broken.
Meter not calibrated.
Wrong temp value.
Check electrode is connected.
Clean or replace electrode.
Recalibrate the meter.
Confirm measurement condition.
Treat samples to bring within
meter measuring range.
Reset meter.
Slow
response
Dirty electrode.
Temperature of sample is
changing.
Clean the electrode.
Allow temperature to stabilize.
Meter not
responding to
key press
Manual HOLD or Auto
HOLD is active.
Worn keypad.
Press
to deactivate manual
HOLD. See Section 6.4 to disable
Auto Hold feature if enabled.
Contact Technical Service.
Invalid key; Button is not
functional in the current
operation mode.
Press alternate key.
Select valid key depending on
current mode.
Secondary
display
continually
scrolls
Conductivity calibration
standard is not within 40%
of expected value during
automatic calibration.
Ensure electrode guard has not
been removed.
Use fresh or different calibration
standard.
16