3.0 INSTALLATION

3.1 LOCATING THE SCALE



• The scales should not be placed in a location that

will reduce the accuracy.

• Avoid extremes of temperature. Do not place in

direct sunlight or near air conditioning vents.

• Avoid unsuitable tables. The table or floor must be

rigid and not vibrate.

• Avoid unstable power sources. Do not use near

large users of electricity such as welding
equipment or large motors.

• Do not place near vibrating machinery.

• Avoid

high

humidity

that

might

cause

condensation. Avoid direct contact with water. Do
not spray or immerse the scales in water.

• Avoid air movement such as from fans or opening

doors. Do not place near open windows or air-
conditioning vents.

• Keep the scales clean. Do not stack material on

the scales when they are not in use.


3.2 INSTALLATION OF CBC SCALES

• The CBC Series come with a stainless steel platform packed

separately.

• Place the platform in the locating holes on the top cover.

Do not press with excessive force as this could damage the load

cell inside.

• Level the scale by adjusting the four feet. The scale should be

adjusted such that the bubble in the spirit level is in the centre of
the level and the scale is supported by all four feet.

• Turn the power ON using the switch located on the right side of

the base.